Emission Control: Canister: Inspection
- INSPECT CANISTER ASSEMBLY
- Visually check the canister for cracks or damage. If cracks or damage are found, replace the canister assembly.
- Check the canister operation.
- With the purge port closed, blow 0.39 kPa (4.0 kgf/cm2 0.06 psi) of air into the vent port, and check that air flows from the air inlet port. If the result is not as specified, replace the canister assembly.
- With the vent port closed, blow 0.39 kPa (4.0 kgf/cm2 0.06 psi) of air inlet port, and check that air flows from the purge port. If the result is not as specified, replace the canister assembly.
- Check the airtightness.
- While holding the vent and air inlet ports closed, apply vacuum (3.43 kPa, 25.7 mmHg, 1.01 in.Hg) to the purge port, and check that the vacuum is maintained for 1 minute. If the result is not as specified, replace the canister assembly.
- Check the leak detection pump.
- Remove the detection pump from the canister.
- Check that air flows from port A to B and then to C.
If the result is not as specified, replace the canister assembly.
- Connect the positive (+) lead of the battery to terminal 7 and the negative (-) lead to terminal 6.
- Check that the valve is closed.
If the result is not as specified, replace the canister assembly.
- Install the detection pump.